- Research
- Open access
- Published:
Classifying physical activity levels using Mean Amplitude Deviation in adults using a chest worn accelerometer: validation of the Vivalink ECG Patch
BMC Sports Science, Medicine and Rehabilitation volumeĀ 16, ArticleĀ number:Ā 212 (2024)
Abstract
Background
The development of readily available wearable accelerometers has enabled clinicians to objectively monitor physical activity (PA) remotely in the community, a superior alternative to patient self-reporting measures. Critical to the value of these monitors is the ability to reliably detect when patients are undergoing ambulatory activity. Previous studies have highlighted the strength of using mean amplitude deviation (MAD) as a universal measure for analysing raw accelerometery data and defining cut-points between sedentary and ambulatory activities. Currently however there is little evidence surrounding the use of chest-worn accelerometers which can provide simultaneous monitoring of other physiological parameters such as heart rate (HR), RR intervals, and Respiratory Rate alongside accelerometery data. We aimed to calibrate the accelerometery function within the VivaLink ECG patch to determine the cut-point MAD value for differentiating sedentary and ambulatory activities.
Methods
We recruited healthy volunteers to undergo a randomised series of 9 activities that simulate typical free-living behaviours, while wearing a VivaLink ECG Patch (Campbell, California). MAD values were applied to a Generalised Linear Mixed Model to determine cut-points between sedentary and ambulatory activities. We constructed a Receiver Operating Characteristic (ROC) curve to analyse the sensitivity and specificity of the cut-off MAD value.
Results
Eighteen healthy adults volunteered to the study and mean MAD values were collected for each activity. The optimal MAD cut-point between sedentary and ambulatory activities was 47.73mG. ROC curve analysis revealed an area under the curve of 0.99 (pā<ā0.001) for this value with a sensitivity and specificity of 98% and 100% respectively.
Conclusion
In conclusion, the MAD cut-point determined in our study is very effective at categorising sedentary and ambulatory activities among healthy adults and may be of use in monitoring PA in the community with minimal burden. It will also be useful for future studies aiming to simultaneously monitor PA with other physiological parameters via chest worn accelerometers.
Background
Accelerometers are electrochemical sensors that detect and quantify acceleration of an object in three planes and have become incorporated into a variety of electronic devices, smartphones, and wrist-watches [1]. These devices have enabled remote monitoring of individualās physical activity levels in the free-living environment over extended periods of time with minimal burden, and thus have developed interest in the fields of sport science, rehabilitation, and many branches of clinical medicine, as they can be applied to a wide range of clinical scenarios and research [2]. The importance of physical activity to health and wellbeing across many physiological systems has been well documented [3] but current methods of measuring these metrics, which rely on self-reporting by patients, suffer the limitations of recollection and response bias and therefore may lack reliability and validity [4, 5]. The rapidly growing market of wearable activity monitors (WAMs) [6], which utilise accelerometers, may offer a cheap and accessible solution by providing an objective measure of physical activity duration and intensity. Recent systematic reviews have already suggested the benefit of utilising WAMs in increasing physical activity in patients with type 2 diabetes [7], cancer [8], and those undergoing cardiac rehabilitation [9], with many other potential clinically significant benefits yet to be explored.
Numerous accelerometery validation studies have been conducted to define specific cut-off points in physical activity intensity levels in adults [10,11,12,13] and children [14], however this is predominantly in wrist and hip worn accelerometers such as GENEActive [15] and ActiGraph [16]. Each accelerometer brand and model has a unique algorithm for converting raw accelerometery data into counts per minute and then using cut-points to classify the intensity of the physical activity, moreover, they are placed in different locations of the body. Therefore, direct comparisons of cut-points for physical activity intensity for one brand cannot be generalised to others.
Currently there is a lack of evidence surrounding the use and calibration of chest-worn accelerometers which have several advantages over wrist worn devices. Chest-worn accelerometers are less likely to detect artefactual limb movement and [17] can allow simultaneous monitoring of other physiological parameters such as heart rate variability, heart rate (HR) and respiratory rate (RR) in a single device [18, 19].Moreover, they hold clinical benefits as analysis of the ECG rhythm can allow diagnosis of conditions such as exercise induced arrhythmia.
Recent studies have demonstrated the benefit of using mean amplitude deviation (MAD) in analysing raw acceleration data to define cut-points between sedentary and ambulatory activities in adults [10] and children [14].Sedentary behaviour is typically defined as any waking behaviour done while lying, reclining, sitting, or standing, with no ambulation and ambulatory activity is considered any behaviour involving participant ambulation typically with a slow-paced walk as the lowest accepted activity intensity. Continuous assessment of MAD allows the userās activity state to be determined at any given point and contemporaneous analysis of physiology during exercise states.
If cut-points in MAD values for defining sedentary and ambulatory behaviour are defined, and these demonstrate high sensitivity and specificity, it will allow further application and study of this accelerometer. Moreover, it will be useful for clinicians aiming to monitor HR, RR, or other parameters offered by the ECG Patch, remotely in free-living conditions in conjunction with accelerometery. The authors for example seek to use this monitor to assess pre-operative fitness by monitoring heart rate recovery in the community setting during activities of daily living, but such validation work is essential and will be useful to other investigators in a range of clinical and non-clinicals scenarios.
The VivaLink ECG patch is a chest worn consumer-orientated accelerometer and is applied to the userās chest with an adhesive strip and can provide simultaneous monitoring of many physiological parameters such as, HR, RR, body temperature, and HR variability. In this study we aim to calibrate the VivaLink ECG Patch (Campbell, USA) [18] accelerometer and define cut-points in accelerometery obtained from healthy volunteers during activities of daily living data so be able to identify when the wearer is ambulatory or sedentary.
Methods
Participants
Participants were recruited by convenience sampling from a publicly distributed email around the study centre. Participants provided informed consent to participate in the study which was approved by the University of Glasgow Medicine, Veterinary and Life Sciences Ethics committee (reference number: 200220144).
Accelerometer
We used the VivaLink ECG Patch VV330 (VivaLink, Campbell, USA), a chest worn HR monitor with a built-in tri-axial accelerometer. The ECG Patch was applied to each participant under the guidance of the investigator to ensure consistency in the deviceās placement (Image 1) according to the manufacturerās instructions.
Procedures
Baseline demographic data was collected from each participant who then undertook a series of supervised 2-minute physical activities which imitate typical free-living behaviours. We aimed to validate our accelerometer by analysing accelerometery data against these activities in a controlled environment. There were 9 activities in total and the order of the activities was randomised for each participant using a computer-generated sequence on Microsoft Excel (Microsoft Corp., Redmond, WA). The 9 activities were classified as either sedentary or ambulatory. The sedentary activities included: lying supine, sitting, standing and remaining stationary, typing at computer while seated, writing while seated and standing while moving a 1 kg hand weight. The ambulatory activities included: slow paced walking, normal paced walking, and brisk paced walking. For the duration of these activities, participants were instructed not to talk or engage in any other tasks. These procedures are used as a standard in accelerometery validation studies [10, 14]. Walking paces were self-determined by the participants and the researcher timed the duration of each activity using a stopwatch. The time each activity began and ended was recorded to the nearest second.
Data processing
Raw accelerometer data was collected in milligravitational (mg) units by the ECG Patch. Accelerometer data for each activity for each participant was stored on a work station and the Mean Amplitude Deviation (MAD) for each activity was analysed in 5 s epochs; 5 s epochs have been established as the optimum epoch length in previous studies [10, 20, 21].
MAD is a statistical measure describing the mean distance of the data point about the mean. This allows quantification and interpretation of accelerometer data, and is calculated as follows:
where n is the number of data points in the epoch, ši is the ith resultant sample from the epoch and šĢ is the resultant value of the epoch [22].
For each activity the mean MAD and HR values were calculated from between 15 s after the start of the activity and 15 s before the end of the activity, to minimise the effect changing between activities has on movement and HR data. The mean MAD values for each activity for each participant would then be used to determine the optimum cut-off value between sedentary and ambulatory activities.
Statistics
MAD values are described as mean and standard deviation. To determine the optimum cut-point in the MAD values derived from accelerometer measurements, we employed a Generalised Linear Mixed Model (GLMMs) approach to perform a logistic regression, by incorporating sedentary and ambulatory behaviours as the binary outcome of interest and the MAD values derived from accelerometer measurements as predictors in our model. To assess the performance of the model and to determine the optimal MAD cut-point, we performed Receiver Operating Characteristic (ROC) curve analysis. The optimal ROC cut-off was defined as the value with the highest sum of specificity and sensitivity (Youdenās index).
To investigate the relationship between MAD and HR, an analysis of covariance (ANCOVA) was conducted. This allowed comparison of the means within each participant while considering the effect of continuous covariates.
Results
Eighteen healthy adults volunteered to take part in the study: nine men and nine women with a median age of 30.8 (+/- 6.4 years) years, with a mean height, weight and BMI of 173.7 cm (SD 11.9), 76.7 kg (16.3), and 25.3 kg/m2 (4.4) respectively. Volunteer demographics are summarised in Table 1; in general, our volunteers reflected a younger and healthier population as recruitment took place in a University Healthcare setting. Table 2 shows the mean MAD value for each subject during each activity and Fig. 1 shows the mean MAD value of all 18 participants for each of the 9 activities over time. Mean MAD values range from 4.1 in the most sedentary activity of lying supine to 319.6 in the most vigorous ambulatory activity of brisk walking. In all cases, ambulatory activities had higher mean MAD values compared to sedentary activities.
Graph representing MAD versus time for each activity - Mean MAD (expressed in milligravity units (mg)) for each participant versus time for each of the 9 different activities with the 95% confidence intervals. Activities were preformed in a randomised order but have been re-ordered for presentation to aid interpretation
The mean MAD value for all sedentary activities was 7.4 compared to the mean MAD value for ambulatory activities of 220.6 (pā=ā0.007). Analysis of ROC curve, (Fig. 2), determined that the optimal cut-point in the MAD values between sedentary and ambulatory behaviour was 47.73 mg. The area under the ROC curve was 0.99 (pā<ā0.001) with a sensitivity and specificity of both 100%.
ANCOVA analysis revealed a correlation coefficient of 0.94 (pā<ā0.001) between HR and MAD values, suggesting a very strong positive correlation between MAD and HR values (Fig. 3). Inspection of Fig. 3 suggests the existence of a consistent linear relationship between MAD and HR.
Discussion
In this study we calibrated the VivaLink ECG Patch, a chest-worn device with a tri-axial accelerometer, to determine the mean MAD values for various activities of free-living. We successfully recruited an equal number of men and women to the study with an average age and BMI lower than the UK average [23,24,25], reflecting the recruitment setting being a University Hospital. Visual inspection of our MAD results reveals a clear delineation between sedentary and ambulatory activities with no overlap of the confidence intervals with ambulatory activities averaging a significantly higher MAD value compared to sedentary activities. This is supported by our MAD cut-off value demonstrating a very high sensitivity and specificity allowing us to accurately and precisely discriminate between the intensity of activities users undertake in the community. Moreover, our ANCOVA analysis of HR and MAD supported the validity of our pattern-based intensity classification and aligns with the results we would anticipate with the greater degree of movement resulting in a higher heart rate. This suggests that this accelerometer could be of benefit in remote monitoring of physical activity levels, alongside other physiological metrics irrespective of sex and body morphology of the user. A previous study assessing HR recovery in the community also utilised the VivaLink ECG Patch and demonstrated that participants were comfortable using the device without supervision of an investigator [26]. This may be useful in developing research exploring the benefits of digital health [27], wearable technology [28] and āHospital at Homeā [29]. There are some limitations to this study to consider. Firstly, our volunteers were a sample of healthy, relatively young adults and as such our findings may not therefore be generalisable to adolescents, children or older adults, particular those who may be faced with frailty. However, a previous study found that similar MAD cut-off values were derived for adolescents and adults when using MAD to quantify raw accelerometery and classify activity intensity [14], highlighting the potential utility of our results to a wider population. Secondly, although the activities the participants undertook imitated those of free-living, they were conducted in a laboratory setting under continuous supervision which may have resulted in altered behaviours and influenced both HR and the dynamics of movement. Participants were also instructed not to talk or use phones during these activities, potentially limiting the accuracy of how well they represent activities of daily living.
No formal reproducibility analysis was conducted comparing the results of each activity across participants. However, we had multiple activities classified as either sedentary or ambulatory being analysed against each other for each participant thus demonstrating some reproducibility in our results. Finally, the 9 activities chosen to be investigated represent only a fraction of the number of potential free-living sedentary and ambulatory activities that could be undertaken, for example, driving, cooking, running, swimming or weight-training. If more activities were assessed, there may be sedentary and ambulatory activities which are harder to discriminate purely using a MAD-based cut-point value. However, the activities we chose are frequently used in similar studies [10, 14] and are common examples used by the Sedentary Behaviour Research Network [30].
Conclusion
The MAD cut-point determined in our study is effective at differentiating between sedentary and ambulatory activities among healthy adults using the VivaLink ECG Patch. This may benefit future studies which utilise chest-worn accelerometers investigating free-living physical activity in conjunction with other physiological parameters.
Data availability
Data are available on reasonable request - deidentified participant data is available from the corresponding author subject to approval of the sponsor organisation.
Abbreviations
- PA:
-
Physical Activity
- HR:
-
Heart Rate
- RR:
-
Respiratory rate
- MAD:
-
Mean Amplitude Deviation
- ROC:
-
Receiver Operating Characteristic
- ECG:
-
Electrocardiogram
- WAM:
-
Wearable Activity Monitor
- mg:
-
Milligravity (the unit for MAD)
- RR intervals:
-
The time elapsed between two successive R waves of the QRS signal on the ECG
References
Kumar K, Sharma A, Tripathi SL. Sensors and their application. Elsevier Inc.; 2021. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/B978-0-323-85172-5.00021-6.
Maher C, Szeto K, Arnold J. The use of accelerometer-based wearable activity monitors in clinical settings: current practice, barriers, enablers, and future opportunities. BMC Health Serv Res. 2021;21:1064. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12913-021-07096-7.
Benefits of Physical Activity. Centers Dis Control Prev 2023. https://www.cdc.gov/physicalactivity/basics/pa-health/index.htm#:~:text=Regular physical activity is one,ability to do everyday activities. (accessed September 20, 2023).
Prince SA, Adamo KB, Hamel M, Hardt J, Connor Gorber S, Tremblay M. A comparison of direct versus self-report measures for assessing physical activity in adults: a systematic review. Int J Behav Nutr Phys Act. 2008;5:56. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/1479-5868-5-56.
Helmerhorst HHJ, Brage S, Warren J, Besson H, Ekelund U. A systematic review of reliability and objective criterion-related validity of physical activity questionnaires. Int J Behav Nutr Phys Act. 2012;9:103. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/1479-5868-9-103.
Laricchia F. Forecast wearables unit shipments worldwide from 2014 to 2024. Statista. 2023. https://www.statista.com/statistics/437871/wearables-worldwide-shipments/ (accessed September 18, 2023).
Baskerville R, Roberts N, Farmer A. Systematic review or Meta-analysis impact of accelerometer and pedometer use on physical activity and glycaemic control in people with type 2 diabetes : a systematic review and meta-analysis 2017:612ā20. https://doiorg.publicaciones.saludcastillayleon.es/10.1111/dme.13331
Schaffer K, Panneerselvam N, Loh KP, Herrmann R. Systematic review of Randomized controlled trials of Exercise interventions using Digital Activity trackers in patients. Cancer. 2019;17:57ā63. https://doiorg.publicaciones.saludcastillayleon.es/10.6004/jnccn.2018.7082.
Hannan AL, Harders MP, Hing W, Climstein M, Coombes JS, Furness J. Impact of wearable physical activity monitoring devices with exercise prescription or advice in the maintenance phase of cardiac rehabilitation. Syst Rev meta-analysis. 2019;8:1ā21.
VƤhƤ-YpyƤ H, Vasankari T, Husu P, Suni J, SievƤnen H. A universal, accurate intensityābased classification of different physical activities using raw data of accelerometer. Clin Physiol Funct Imaging. 2015;35:64ā70. https://doiorg.publicaciones.saludcastillayleon.es/10.1111/cpf.12127.
Holmlund T, Ekblom-bak E, FranzĆ©n E, Hultling C, Wahman K. Defining accelerometer cut-points for different intensity levels in motor-complete spinal cord injury. Spinal Cord 2020:116ā24. https://doiorg.publicaciones.saludcastillayleon.es/10.1038/s41393-019-0308-y
ReportLinker. The global accelerometer medical sensors market 2023.
Migueles JH, Cadenas-Sanchez C, Alcantara JMA, Leal-MartĆn J, MaƱas A, Ara I, et al. Calibration and cross-validation of accelerometer cut-points to classify Sedentary Time and Physical Activity from Hip and Non-dominant and Dominant wrists in older adults. Sensors. 2021;21:3326. https://doiorg.publicaciones.saludcastillayleon.es/10.3390/s21103326.
Aittasalo M, VƤhƤ-YpyƤ H, Vasankari T, Husu P, Jussila A-M, SievƤnen H. Mean amplitude deviation calculated from raw acceleration data: a novel method for classifying the intensity of adolescentsā physical activity irrespective of accelerometer brand. BMC Sports Sci Med Rehabil. 2015;7:18. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13102-015-0010-0.
GENEActiv: Raw Data Accelerometry. Activinsights 2023. https://activinsights.com/technology/geneactiv/ (accessed September 18, 2023).
ActiGraph wGT3X-BT. ActiGraph 2023. https://theactigraph.com/actigraph-wgt3x-bt (accessed September 18, 2023).
Cleland I, Kikhia B, Nugent C, Boytsov A, Hallberg J, Synnes K, et al. Optimal Placement of Accelerometers for the detection of everyday activities. Sensors. 2013;13:9183ā200. https://doiorg.publicaciones.saludcastillayleon.es/10.3390/s130709183.
Wearable Heart and ECG Monitor Patch. VivaLink Co 2023. https://www.vivalink.com/wearable-ecg-monitor (accessed January 12, 2023).
GARMIN, HRM-Pro⢠P. 2023. https://www.garmin.com/en-GB/c/wearables-smartwatches/ (accessed September 20, 2023).
Migueles JH, Cadenas-Sanchez C, Rowlands AV, Henriksson P, Shiroma EJ, Acosta FM, et al. Comparability of accelerometer signal aggregation metrics across placements and dominant wrist cut points for the assessment of physical activity in adults. Sci Rep. 2019;9:18235. https://doiorg.publicaciones.saludcastillayleon.es/10.1038/s41598-019-54267-y.
Matthews CE, Hagstrƶmer M, Pober DM, Bowles HR. Best practices for using physical activity monitors in Population-Based Research. Med Sci Sport Exerc. 2012;44:S68ā76. https://doiorg.publicaciones.saludcastillayleon.es/10.1249/MSS.0b013e3182399e5b.
Bakrania K, Yates T, Rowlands AV, Esliger DW, Bunnewell S, Sanders J, et al. Intensity thresholds on raw acceleration data: Euclidean Norm Minus one (ENMO) and Mean Amplitude deviation (MAD) approaches. PLoS ONE. 2016;11:e0164045. https://doiorg.publicaciones.saludcastillayleon.es/10.1371/journal.pone.0164045.
Mean body mass. index (BMI) of adults in England 2021, by gender and age. Statista 2021. https://www.statista.com/statistics/375886/adult-s-body-mass-index-by-gender-and-age-in-england/ (accessed August 20, 2024).
Mean body mass index. of adults in Scotland in 2022, by gender and age. Statista 2021. https://www.statista.com/statistics/374006/adult-mean-body-mass-index-by-gender-and-age-in-scotland-uk/#:~:text=This statistic displays the mean,30 is classified as overweight. (accessed August 20, 2024).
Median age of the population of the United Kingdom. from 2001 to 2022. Statista 2022. https://www.statista.com/statistics/281288/median-age-of-the-population-of-the-uk/#:~:text=In 2022%2Ā C the median age,median age remained at 40. (accessed August 20, 2024).
Luckhurst J, Hughes C, Shelley B. Adding objectivity to submaximal exercise testing by assessment of heart rate recovery at homeāa healthy volunteer study iv (search-iv). Br J Anaesth. 2023;131:e87ā8. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.bja.2023.06.007.
Digital Health. World Heal Organ n.d. https://www.who.int/health-topics/digital-health#tab=tab_3 (accessed November 8, 2022).
The Wearable Health Tech Revolution. Healthline 2023. https://transform.healthline.com/future-of-health/the-wearable-health-tech-revolution (accessed September 20, 2023).
Hospital at Home. Heal Improv Scotl 2023. https://ihub.scot/project-toolkits/frailty-improvement-programme/frailty-resources/hospital-at-home/ (accessed September 13, 2023).
SBRN Terminology Consensus Project. Sedentary Behav Res Netw 2020. https://www.sedentarybehaviour.org/sbrn-terminology-consensus-project/#consensus-definitions (accessed October 4, 2023).
Acknowledgements
The authors would like to acknowledge Dr Martin Shaw for his advice on the statistical analysis of the data.
Funding
This study was grant funded by The Scottish Society of Anaesthetists and Medical Research Scotland. VivaLink have contributed to open access publishing costs. Neither party was involved in the studyās design, data analysis or conclusions nor exerted any editorial control on the manuscript.
Author information
Authors and Affiliations
Contributions
J.L. wrote the main manuscript text and prepared all figures, tables, and study design. B.S. reviewed the manuscript, figures, tables and both B.S. and C.H. provided supported in the design of the study.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
All participants of the study provided written informed consent. Ethical approval was granted by the University of Glasgow Medicine Veterinary and Life Sciences Ethics Committee (reference number: 200220144) .
Consent for publication
Not applicable.
Competing interests
The APC fee is being paid for by VivaLink, the company who provided the wearable heart rate monitor for this study.
Additional information
Publisherās note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the articleās Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the articleās Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Luckhurst, J., Hughes, C. & Shelley, B. Classifying physical activity levels using Mean Amplitude Deviation in adults using a chest worn accelerometer: validation of the Vivalink ECG Patch. BMC Sports Sci Med Rehabil 16, 212 (2024). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13102-024-00991-6
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13102-024-00991-6