Reproducibility of peak force for isometric and isokinetic multi-joint leg extension exercise

Abstract

Background

Isokinetic dynamometry is a common tool for evaluating muscle function and is used across various disciplines. Technical advancements have shifted focus towards multi-joint exercises such as the leg press, offering insights into practical human movement dynamics. However, previous reproducibility studies have focused predominantly on single-joint exercises, warranting investigations into the reliability of multi-joint exercises. This study aimed to assess the reproducibility of peak force (PF) during multi-joint leg press exercises using the IsoMed 2000 dynamometer.

Methods

Thirty physically active subjects (mean: stature 179.4 cm; body mass 76.0 kg; age 30.6 years) participated in three testing sessions. Each session consisted of isometric and isokinetic leg press exercises. Knee angles for isometric exercises included 100° and 140°; velocities for isokinetic exercise included 30 mm/s and 600 mm/s. The first session served as the familiarization session. Statistical analysis included paired sample t-tests, Cohen’s d effect sizes, intraclass correlation coefficients (ICC), standard errors of measurement (SEM), and Bland-Altman calculations, including corresponding plots.

Results

Descriptive data revealed consistent PF across sessions, with a significant between-sessions difference observed only for isometric (100°) leg extension in the right leg (p < 0.001; d = 0.13). ICC calculations showed high relative reproducibility (ICC > 0.911), with SEM ranging from 37.6 to 294.7 N (SEM% 2.3–6.3%, respectively). Bland-Altman plots depicted minimal intersession disparities (-141.8–68.3 N, respectively − 3.02–1.26%), supporting high reliability.

Conclusions

This study highlights the reliability of assessing peak force during isometric and isokinetic leg press exercises using the IsoMed 2000 after a single familiarization session. These findings support its utility in muscular performance evaluation, urging practitioners to incorporate familiarization trials for accurate assessments.

Peer Review reports

Introduction

The introduction of the isokinetic concept in the 1960s [1] marked a paradigm shift in muscle function assessment, with isokinetic dynamometry establishing itself as a widely accepted method for evaluating muscular performance [2,3,4]. Nowadays, isokinetic exercise, which is characterized by constant velocity movements against modulated resistance, has become an integral part of performance evaluation and training. Typical applications encompass a spectrum of disciplines within sport and medicine, including orthopedics, performance sports, rehabilitation, and sports physical therapy.

In the past, a majority of studies executing isokinetic dynamometry used single-joint exercises [5]. However, due to the technical development of multi-joint testing devices, there has been increasing interest in leg press exercises in recent years. The use of such devices enables the evaluation of muscular function within a multi-joint exercise, which is suggested to have greater practical relevance for actual everyday as well as sport-specific human movements than a single-joint exercise [6, 7]. For example, significant correlations were found for isokinetic leg press and several jumping exercises, as well as for squat and sprint performance [8,9,10]. The isokinetic multi-joint leg press exercise has therefore already been used for a variety of purposes, including the evaluation of different training and rehabilitation interventions [11,12,13,14,15,16,17]. In all of these settings, reliable data on muscular performance are paramount for optimizing training strategies, monitoring rehabilitation progress, and identifying potential performance limitations.

Previous studies on the topic of dynamometric reproducibility have focused mainly on single-joint exercises [5, 18,19,20,21,22], with only a handful investigating multi-joint leg press [23,24,25,26].

However, reproducibility results from a single-joint exercise cannot be transferred to a multi-joint exercise such as a leg press. It cannot be assumed that the generally high reliability of the knee joint automatically applies to other joints [27]. This is supported by the fact that only weak and non-significant correlations between single-joint and multi-joint isokinetic tests of the hip and knee extensors have been found [16, 28]. In general, reproducibility data should always be considered specific to the respective measurement device, test protocol, and subject group.

To the best of the authors’ knowledge, there is currently only one study examining the reproducibility of multi-joint leg extension exercises using the IsoMed 2000 device [23]. However, the authors used a different protocol in their study, investigating only moderate isokinetic velocities and no isometric contractions.

The aim of this study was, therefore, to assess the reproducibility of peak force (PF) for isometric and isokinetic multi-joint leg press exercise at two different isokinetic movement velocities (30 mm/s and 600 m/s) and at two different isometric knee angles (100° and 140°), using the IsoMed 2000 dynamometer.

Methods

Subjects

Thirty physically active subjects (25 male, 5 female; mean (SD): stature 179.4 (8.4) cm; body mass 76.0 (9.9) kg; age 30.6 (8.2) years) volunteered to participate in this study. An inclusion criterion for this study was the absence of any major previous orthopaedic lower extremity pathologies that would have needed clinical treatment. All subjects were physically active on a recreational level but did not have any previous experience in isokinetic exercise. Before starting the experiments, all participants received instructions to arrive at the laboratory in a mentally and physically rested state. Subjects were asked to ingest their last main meal at least 3 h before each test, to avoid any consumption of caffeine for 12 h, and to not engage in vigorous physical activity for 48 h before each test session. Written informed consent was provided by all subjects, and they were advised that withdrawal from the study is possible at any time. All subjects were fully informed about the experimental procedures and elucidated about the risks and benefits of participating in the study. The local research ethics board at the University of Applied Sciences Wiener Neustadt approved this study on the 5th of April 2021 (approval nr RB20210405013). All experiments conformed to the Declaration of Helsinki [29].

Instruments

All experiments were conducted on an IsoMed 2000 dynamometer (D. & R. Ferstl GmbH, Hemau, Germany). The manufacturer provides different adaptors for this device, including a so-called athletic linear module that was used for this study. This module enables the conversion of the dynamometer into a motor-driven leg press (Fig. 1). A drive shaft (a) is used for converting the rotational motion of the dynamometer into a translational motion at the leg press via a tooth-belt linear drive. This drive shaft offers two gears (b) that differ in their ratio (gear I = 1:1; gear II = 1:1.5) and therefore determines the maximum linear movement speed (800 mm/s vs. 1,200 mm/s) and the maximum force (8,850 N vs. 5,850 N). Strain gauge force sensors (c) that are located at the back of the footrest (d) are used for the measurement of the applied force separately for the left and right legs.

For the present study, gear I of the drive shaft was used. Before each session, the device was calibrated according to the manufacturer’s instructions. Data recording was performed at a sample rate of 200 Hz using the manufacturer’s integrated computer software IsoMed Analyse SP3-i51.

Fig. 1
figure 1

The IsoMed 2000 device with athletic linear module and a subject in the starting position. The IsoMed 2000 basic device converted to a motor-driven leg press using the manufacturers athletic linear module: a: drive shaft; b: selectable gears; c: force sensors; d: footrest

Procedures

Subjects were tested in three identical sessions. The first session was used as a familiarization trial to accommodate participants to the isokinetic device and protocol, as recommended in several studies [30,31,32,33]. Therefore, data from this session were not included in subsequent analyses. The sessions were typically scheduled 72–96 h apart to permit sufficient recovery for the subjects. Sessions were conducted at the same time of day (± 1.5 h) for each subject to minimize possible influences from diurnal variations. In an effort to minimize possible inter-tester variability, all tests in this study were conducted by the same examiner, who is a trained sport- and exercise scientist. Subjects were asked to arrive at the laboratory with shoes that had a hard rubber outsole without additional cushioning elements and to wear the same shoes for all sessions.

Each session started with a 10-min general warm-up, followed by the initial positioning of the subject on the device. The vertical backrest of the dynamometer was reclined to 75°, and the footrest was set at an inclination of 10° towards plantar flexion of the ankle. For fixation of the subject to the device and to minimize errant body movements, adjustable seat belts and pads were firmly applied, including the shoulders, thorax, and hip. In addition, the subjects were instructed to grasp the side handles of the device with their hands.

The range of motion and velocity served as the basic input parameters [34]. A mechanical handheld goniometer was utilized to set the range of motion to 90–170° (180° = fully extended). To do so, the trochanter major, lateral femoral epicondyle, and lateral malleolus were used as bony reference points and detected by palpation. The start and end of movement were measured while contracting the implicated muscles. For safety reasons and to minimize any risk of injuries, the popliteal pad of the device was adjusted and placed under the popliteal fossa of the subject in a way that prevented a complete extension (or even hyperextension) of the knee joint.

The subjects were then asked to put the leg onto the leg press footrest so that the heel ended up on the heel support area of the device. For isometric measurements, the footrest was fixed to achieve knee joint angles of 100° (Iso100) and 140° (Iso140). The position of the footrest used to achieve these angles was individually adjusted for each subject using the goniometer and the methods mentioned before.

Integrated software was used to record individual settings for each subject to guarantee identical positioning in every session. In each session, subjects performed isometric and isokinetic leg extensions in two conditions each. The conditions included isometric measurements with knee angles of 100° (Iso100) and 140° (Iso140), and isokinetic measurements that were conducted at translational velocities of 30 mm/s and 600 mm/s. All subjects performed each condition in three different ways: left-leg only (L), right-leg only (R), and two-legged (T). Subjects always started with T, followed by both one-legged conditions. Starting the one-legged conditions with the left or right leg was randomly assigned, although starting with the left vs. starting with the right leg was evenly distributed within the subject group.

The test order was Iso100, followed by Iso140 for isometric conditions. Isokinetic conditions were set at 30 mm/s, followed by 600 mm/s [35, 36]. The test order remained unchanged throughout all sessions. The initial starting position was achieved passively, and all isokinetic measurements were completed as discrete movements in a single direction [37]. To prepare for each condition of the test, subjects engaged in a submaximal specific warm-up on the device to familiarize themselves with the demands of each assessment. This specific warm-up comprised 10 repetitions at an intensity approximating 50% of the maximum voluntary contraction, followed by 3 repetitions at an intensity approximating 80% of the maximum voluntary contraction. Upon completion of this warm-up, subjects received a 3-min break that was used to repeat the explanation of the procedures for the subsequent condition using standardized instructions. Participants performed a minimum of three repetitions for each test condition. However, additional repetitions were performed until PF started to decline. Within a maximum of five repetitions, PF was achieved in all subjects. Before each repetition, participants received 3-min of passive rest to ensure sufficient recovery. Strong verbal encouragement from the examiner and visual feedback on a screen were provided to maximize the effort of the subjects.

Statistical analysis

Two subjects failed to complete the entirety of the scheduled sessions. Consequently, their data were excluded, and subsequent analytical procedures were conducted on the remaining dataset, comprising 28 subjects. The main output parameter that was used for further analysis was the PF, as this parameter is commonly used in numerous studies dealing with muscular performance and dynamometers [34, 38]. For each condition, the repetition with the highest PF value was extracted from each session and subjected to analysis. Descriptive data are presented as means (± standard deviation). The normality of the data distribution was assessed using the Shapiro-Wilk test. To assess possible differences between PF measurements, paired sample t-tests were applied. The magnitude of the difference between sessions was calculated using Cohen’s d effect sizes, applying the formula: \(\:\left(Mea{n}_{Session2}-Mea{n}_{Session3}\right)/SDpooled\). These were interpreted according to Cohen [39], where|d|0.2 = small,|d|0.5 = moderate, and|d|0.8 = large. Relative reproducibility was evaluated using the two-way random effect intraclass correlation coefficient (ICC) and interpreted by Vincent’s criteria [40]. According to these guidelines, an ICC exceeding 0.9 is considered high, while values between 0.8 and 0.9 are deemed moderate, and those below 0.8 are considered low. Absolute reproducibility was evaluated by calculating the standard error of measurement (SEM) using the formula \(\:SEM=SD\:\times\:\:\sqrt{\:1-ICC}\) [41, 42]. Additionally, the SEM% was computed, defined as SEM/(means of measurements from session) * 100. To quantify the level of agreement between sessions, Bland-Altman statistics ± 95% limits of agreement (LoA) were calculated, and corresponding plots were generated to visually depict individual results [43]. Statistical analyses were conducted employing IBM SPSS Statistics for Windows, version 28.0 (IBM Corp., Armonk, NY, USA). Figures were created using GraphPad Prism V.9.3 for Windows (GraphPad Software, San Diego, CA, USA). The threshold for statistical significance was set at p < 0.05.

Results

PF was significantly higher (+ 44.2 N) during the third session for Iso100 in the right leg condition (t(27)= -3.69, p = < 0.001; d = 0.13). No other significant differences in PF between sessions were observed. For comparison of sessions, small effect sizes in the range of d < 0.01 to 0.14 were found (Table 1).

The ICC results indicate high relative reproducibility, with values ranging from 0.911 to 0.989 (95% CI 0.809 to 0.995). The absolute reproducibility was expressed as SEM and SEM% and revealed values from 37.6 to 294.7 N and 2.3 to 6.3%, respectively (Table 2).

Bland-Altman plots illustrate a stochastic relationship for isometric and isokinetic leg extension between subject-specific differences and session averages (Figs. 2 and 3). The bias, which represents the average difference between sessions (Table 3), ranged from − 141.8 to 68.3 N (95% LoA from − 1246.0 to 962.4), where a negative value indicates that session three had a greater value than session two.

Table 1 Mean (SD) for peak force measurements as well as p-values and cohens’ d effect sizes for comparison of sessions
Table 2 Relative and absolute reproducibility statistics for comparison of peak force from sessions 2 and 3
Table 3 Bias (SD) in Newton and % as well as 95% LoA for comparison of sessions 2 and 3 received from Bland-Altman calculations
Fig. 2
figure 2

Bland-Altman plots– differences between isometric session two and session three plotted against the means. Differences between session two and session three plotted against the means of session two and session three for (A) two-legged isometric extension at 100° knee angle; (B) left leg only isometric extension at 100° knee angle; (C) right leg only isometric extension at 100° knee angle; (D) two-legged isometric extension at 140° knee angle; (E) left leg only isometric extension at 140° knee-angle; (F) right leg only isometric extension at 140° knee angle

Fig. 3
figure 3

Bland-Altman plots– differences between isokinetic session two and session three plotted against the means. Differences between session two and session three plotted against the means of session two and session three for (G) two-legged isokinetic extension at 30 mm/s; (H) left leg only isokinetic extension at 30 mm/s; (I) right leg only isokinetic extension at 30 mm/s; (J) two-legged isokinetic extension at 600 mm/s; (K) left leg only isokinetic extension at 600 mm/s; (L) right leg only isokinetic extension at 600 mm/s

Discussion and implications

This study aimed to determine the reproducibility of PF during maximum isometric and isokinetic multi-joint leg extension exercises using the IsoMed 2000 dynamometer with a corresponding athletic linear module. In general, the results of this study indicate a high reliability of PF after initial familiarization.

It is a characteristic of isokinetic leg press exercise using a dynamometer such as the IsoMed 2000 that a constant linear velocity of the device’s footrest does not result in constant joint angle velocity throughout the range of motion. In addition, the same linear velocity can result in different angular velocities for subjects with different statures and/or segment lengths [34]. In an attempt to overcome these limitations, a correction procedure has been developed [6]. However, this mathematical model and the underlying assumptions are based on simplified principles that, in practice, only apply to a limited extent. A recent study [44] showed that this procedure is not free of errors and can significantly differ from marker-derived metrics. Due to these concerns and the fact that we controlled for all other possible input parameters (e.g., range of motion, exact placement of subjects), we decided not to apply the correction procedure and instead opted for constant linear velocity throughout all subjects.

Regarding inter-session disparity, no significant differences were found between sessions two and three, except for Iso100 in the right leg. With Cohen’s d effect sizes in the range of < 0.01 to 0.14, all of these differences appear to be small.

Concerning reproducibility, we observed ICC values > 0.911 with a 95% CI of 0.809 or greater. The lowest ICCs were found for slow isokinetic velocities. However, as all ICCs are > 0.9, the relative reproducibility can generally be rated as high for all conditions, in accordance with the recommendations of Vincent [40]. Corresponding SEM measurements of 37.6–294.7 N, or 2.3–6.3%, can be considered appropriate for the majority of common practical applications.

Comparing our ICC results to those of other studies, these results are similar [24,25,26, 45] or greater [46, 47] than those of previous investigations on peak force reproducibility in multi-joint leg extension. However, all these studies used a different device than the one used in our study. The only experiment conducted using the IsoMed 2000 [23] revealed similar results for the ICCs when comparing session two and session three. However, Dirnberger, Huber [23] used a slightly different methodological approach and tested only one leg.

With regard to SEM% results, we observed 2.3 to 6.3% which is lower than in the study of Müller, Baur [46], who found distinctly higher values, reaching 9.5 to 10%, although the authors did not implement a familiarization trial in their study.

In general, our results reveal lower reliability (absolute and relative) for the slower isokinetic velocity in comparison to the faster one. This trend for lower reliability at slower velocities is in agreement with the results of Davies and Heiderscheit [24] and Levine, Klein [48]. According to Levine, Klein [48] the longer contraction time and higher force at slower velocities could lead to a slightly altered position of the subjects on the device. To address this, in our study, we opted to minimize errant body movements using straps and pads. Because of this and that there was only a minimum difference in the magnitude of ICC and SEM between slow and fast velocities, our data suggest that velocity did not affect the results of our study.

The systematic difference, or bias, representing the average disparity between sessions varied from − 141.8 to 68.3 N. Notably, a negative value signifies that session three exhibited a higher average value than did session two. In our study, nine out of twelve conditions revealed a negative bias, illustrating the importance of at least one familiarization trial. However, considering the relative bias (-3.02 to 1.26%), this finding supports the high reliability of the measurement protocol in this study.

Conclusion

This study assessed the reliability of PF measured during various isometric and isokinetic leg press extension exercises, utilizing the IsoMed 2000 dynamometer. The results indicate that using this methodology, PF measurements achieve a level of reliability deemed excellent for most common practical applications. Therefore, professionals are encouraged to incorporate at least one familiarization trial for each chosen angle and velocity when implementing isometric or isokinetic testing, ensuring the validity of subsequent results.

As the isokinetic leg press strength already showed correlations with jump-, squat-, and sprint performance in previous studies and is frequently used for training and rehabilitation interventions, our results could provide valuable information for practitioners when interpreting diagnostic data.

Data availability

The datasets used and analysed during the current study are available from the corresponding author upon reasonable request.

Abbreviations

ICC:

Intraclass correlation coefficient

Iso100:

Isometric 100° knee angle

Iso140:

Isometric 140° knee angle

L:

Left-leg only

LoA:

Limits of agreement

N:

Newton

PF:

Peak force

R:

Right-leg only

SEM:

Standard error of measurement

T:

Two-legged

References

  1. Hislop HJ, Perrine J. The isokinetic concept of exercise. Phys Ther. 1967;47(1):114–7.

    Article  CAS  PubMed  Google Scholar 

  2. Abernethy P, Wilson G, Logan P. Strength and power assessment: issues, controversies and challenges. Sports Med. 1995;19:401–17.

    Article  CAS  PubMed  Google Scholar 

  3. Knapik J, Wright J, Mawdsley R, Braun J. Isokinetic, isometric and isotonic strength relationships. Arch Phys Med Rehabil. 1983;64(2):77–80.

    CAS  PubMed  Google Scholar 

  4. Ly LP, Handelsman DJ. Muscle strength and ageing: methodological aspects of isokinetic dynamometry and androgen administration. Clin Exp Pharmacol Physiol. 2002;29(1–2):39–47.

    Article  CAS  Google Scholar 

  5. Caruso JF, Brown LE, Tufano JJ. The reproducibility of isokinetic dynamometry data. Isokinet Exerc Sci. 2012;20(4):239–53.

    Article  Google Scholar 

  6. Hahn D, Schwirtz A, Huber A. Anthropometric standardisation of multiarticular leg extension movements: a theoretical study. Isokinet Exerc Sci. 2005;13(2):95–101.

    Article  Google Scholar 

  7. Hahn D, Seiberl W, Schmidt S, Schweizer K, Schwirtz A. Evidence of residual force enhancement for multi-joint leg extension. J Biomech. 2010;43(8):1503–8.

    Article  PubMed  Google Scholar 

  8. Manske RC, Smith BS, Rogers ME, Wyatt FB. Closed kinetic chain (linear) isokinetic testing: relationships to functional testing. Isokinet Exerc Sci. 2003;11(3):171–9.

    Article  Google Scholar 

  9. Möck S, Wirth K. Relationship of isokinetic leg press strength to sprinting performance in junior elite volleyball players. German J Sports Med. 2019;70(9):203–8.

    Article  Google Scholar 

  10. Möck S, Erlewein S, Mickel C, Wirth K. Relationship of isokinetic leg press force with jumping performance and 1rm in the squat. J Sci Sport Exerc. 2022;4(2):180–7.

  11. Müller E, Gimpl M, Kirchner S, Kröll J, Jahnel R, Niebauer J, et al. Salzburg skiing for the elderly study: influence of alpine skiing on aerobic capacity, strength, power, and balance. Scand J Med Sci Sports. 2011;21:9–22.

    Article  PubMed  Google Scholar 

  12. Kemmler W, Teschler M, Weißenfels A, Bebenek M, Fröhlich M, Kohl M, et al. Effects of whole-body electromyostimulation versus high-intensity resistance exercise on body composition and strength: a randomized controlled study. Evid Based Complement Alternat Med. 2016;2016:1–9.

    Article  Google Scholar 

  13. Fagnani F, Giombini A, Di Cesare A, Pigozzi F, Di Salvo V. The effects of a whole-body vibration program on muscle performance and flexibility in female athletes. Am J Phys Med Rehabil. 2006;85(12):956–62.

    Article  PubMed  Google Scholar 

  14. Baur H, Müller S, Hirschmüller A, Huber G, Mayer F. Reactivity, stability, and strength performance capacity in motor sports. Br J Sports Med. 2006;40(11):906–11.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Christou EA. Patellar taping increases vastus medialis oblique activity in the presence of patellofemoral pain. J Electromyogr Kinesiol. 2004;14(4):495–504.

    Article  PubMed  Google Scholar 

  16. Feiring DC, Ellenbecker TS. Single versus multiple joint isokinetic testing with ACL reconstructed patients. Isokinet Exerc Sci. 1996;6(2):109–15.

    Article  Google Scholar 

  17. Brown AB, McCartney N, Sale D. Positive adaptations to weight-lifting training in the elderly. J Appl Physiol (1985). 1990;69(5):1725–33.

    Article  CAS  PubMed  Google Scholar 

  18. Adsuar JC, Olivares PR, del Pozo-Cruz B, Parraca JA, Gusi N. Test-retest reliability of isometric and isokinetic knee extension and flexion in patients with fibromyalgia: evaluation of the smallest real difference. Arch Phys Med Rehabil. 2011;92(10):1646–51.

    Article  PubMed  Google Scholar 

  19. Brunner-Althaus C, de Bruin ED. Die Zuverlassigkeit Des Isokinetischen Knie kurzprotokolls Von Swiss Olympic. Schweiz Z Sportmed Sporttraumatol. 2006;54(3):96–100.

    Google Scholar 

  20. Collado-Mateo D, Dominguez-Muñoz FJ, Charrua Z, Adsuar JC, Batalha N, Merellano-Navarro E, et al. Isokinetic strength in peritoneal dialysis patients: a reliability study. Appl Sci. 2019;9(17):3542.

    Article  Google Scholar 

  21. Lund H, Søndergaard K, Zachariassen T, Christensen R, Bülow P, Henriksen M, et al. Learning effect of isokinetic measurements in healthy subjects, and reliability and comparability of Biodex and Lido dynamometers. Clin Physiol Funct Imaging. 2005;25(2):75–82.

    Article  CAS  PubMed  Google Scholar 

  22. Sole G, Hamrén J, Milosavljevic S, Nicholson H, Sullivan SJ. Test-retest reliability of isokinetic knee extension and flexion. Arch Phys Med Rehabil. 2007;88(5):626–31.

    Article  PubMed  Google Scholar 

  23. Dirnberger J, Huber C, Hoop D, Kösters A, Müller E. Reproducibility of concentric and eccentric isokinetic multi-joint leg extension measurements using the IsoMed 2000-system. Isokinet Exerc Sci. 2013;21(3):195–202.

    Article  Google Scholar 

  24. Davies GJ, Heiderscheit BC. Reliability of the Lido Linea closed kinetic chain isokinetic dynamometer. J Orthop Sports Phys Ther. 1997;25(2):133–6.

    Article  CAS  PubMed  Google Scholar 

  25. Papadopoulos C, Theodosiou K, Noussios G, Gantiraga E, Meliggas K, Sambanis M, et al. Evidence for validity and reliability of multiarticular leg extension machine. Int J Appl SciTechnol. 2012;2(8):10–9.

    Google Scholar 

  26. Baur H, Groppa AS, Limacher R, Radlinger L. Low-budget instrumentation of a conventional leg press to measure reliable isometric-strength capacity. J Sport Rehabil. 2016;25(1):1–3.

    Article  Google Scholar 

  27. Wennerberg D. Reliability of an isokinetic dorsiflexion and plantar flexion apparatus. Am J Sports Med. 1991;19(5):519–22.

    Article  CAS  PubMed  Google Scholar 

  28. Dvir Z. An isokinetic study of combined activity of the hip and knee extensors. Clin Biomech Elsevier Ltd. 1996;11(3):135–8.

    Article  CAS  Google Scholar 

  29. Association WM. World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. JAMA. 2013;310(20):2191–4.

    Article  Google Scholar 

  30. Nugent EP, Snodgrass SJ, Callister R. The effect of velocity and familiarisation on the reproducibility of isokinetic dynamometry. Isokinet Exerc Sci. 2015;23(3):205–14.

    Article  Google Scholar 

  31. Roth R, Donath L, Kurz E, Zahner L, Faude O. Absolute and relative reliability of isokinetic and isometric trunk strength testing using the IsoMed-2000 dynamometer. Phys Ther Sport. 2017;24:26–31.

    Article  PubMed  Google Scholar 

  32. Jenkins ND, Cramer JT. Reliability and minimum detectable change for common clinical physical function tests in sarcopenic men and women. J Am Geriatr Soc. 2017;65(4):839–46.

    Article  PubMed  Google Scholar 

  33. Dirnberger J, Wiesinger H-P, Kösters A, Müller E. Reproducibility for isometric and isokinetic maximum knee flexion and extension measurements using the IsoMed 2000-dynamometer. Isokinet Exerc Sci. 2012;20(3):149–53.

    Article  Google Scholar 

  34. Dvir Z, Müller S. Multiple-joint isokinetic dynamometry: a critical review. J Strength Cond Res. 2020;34(2):587–601.

    Article  PubMed  Google Scholar 

  35. Perrin DH. Isokinetic exercise and assessment. Champaign, Il: Human Kinetics; 1993.

    Google Scholar 

  36. Wilhite MR, Cohen ER, Wilhite SC. Reliability of concentric and eccentric measurements of quadriceps performance using the KIN-COM dynamometer: the effect of testing order for three different speeds. J Orthop Sports Phys Ther. 1992;15(4):175–82.

    Article  CAS  PubMed  Google Scholar 

  37. Rothstein JM, Lamb RL, Mayhew TP. Clinical uses of isokinetic measurements: critical issues. Phys Ther. 1987;67(12):1840–4.

    Article  CAS  PubMed  Google Scholar 

  38. Dvir Z. Relevant, less relevant and irrelevant isokinetic strength test parameters: some critical comments. Mov Sport Sciences-Science Motricité. 2013;85(3):15–21.

    Google Scholar 

  39. Cohen J. Statistical power analysis for the behavioral sciences. Second ed. Hillsdale, N.J: Lawrence Erlbaum Associates; 1988.

    Google Scholar 

  40. Vincent WJ. Statistics in Kinesiology. Third ed. Champaign, IL: Human Kinetics; 2005.

    Google Scholar 

  41. Atkinson G, Nevill AM. Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Med. 1998;26:217–38.

    Article  CAS  PubMed  Google Scholar 

  42. Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19(1):231–40.

    PubMed  Google Scholar 

  43. Bland JM, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327(8476):307–10.

    Article  Google Scholar 

  44. Möck S, Günther C, Wirth K. The influence of the calculation method on knee angle trajectories and angle-specific force in multi-joint isokinetic leg extensions. German J Sports Med. 2022;73(4):154–9.

    Google Scholar 

  45. McCarthy CJ, Callaghan MJ, Oldham JA. The reliability of isometric strength and fatigue measures in patients with knee osteoarthritis. Man Ther. 2008;13(2):159–64.

    Article  PubMed  Google Scholar 

  46. Müller S, Baur H, König T, Hirschmüller A, Mayer F. Reproducibility of isokinetic single-and multi-joint strength measurements in healthy and injured athletes. Isokinet Exerc Sci. 2007;15(4):295–302.

    Article  Google Scholar 

  47. Sleivert GG, Wenger HA. Reliability of measuring isometric and isokinetic peak torque, rate of torque development, integrated electromyography, and tibial nerve conduction velocity. Arch Phys Med Rehabil. 1994;75(12):1315–21.

    Article  CAS  PubMed  Google Scholar 

  48. Levine D, Klein A, Morrissey M. Reliability of isokinetic concentric closed kinematic chain testing of the hip and knee extensors. Isokinet Exerc Sci. 1991;1(3):146–52.

    Article  Google Scholar 

Download references

Acknowledgements

The authors express their sincere gratitude to all participants who willingly volunteered their time and effort to contribute to this study.

Funding

This work was supported by Gesellschaft für Forschungsförderung Niederösterreich m.b.H. under grant number SC19-002.

Author information

Authors and Affiliations

Authors

Contributions

KW and MZ conceived and designed the research. MZ conducted the experiments. MZ analysed the data. KW and MZ interpreted the results of the experiments. MZ drafted the first version of the manuscript. AB, AN, KW, and MZ were involved in the revision and approval of the final version of the manuscript. AB supervised the study and was assisted by AN and KW as co-supervisors.

Corresponding author

Correspondence to Manfred Zöger.

Ethics declarations

Ethics approval and consent to participate

The experiments involving human participants were reviewed and approved by the local Research Ethics Board at the University of Applied Sciences Wiener Neustadt on the 5th of April 2021 (approval nr RB20210405013). Written informed consent to participate in this study was provided by all participants before the start of the experiments.

Consent for publication

Written consent for publication has been obtained from the subject in Fig. 1.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zöger, M., Nimmerichter, A., Baca, A. et al. Reproducibility of peak force for isometric and isokinetic multi-joint leg extension exercise. BMC Sports Sci Med Rehabil 17, 14 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13102-025-01061-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13102-025-01061-1

Keywords